Effects of Core Dimensions and Manufacturing Procedure on Fluxgate Noise
Type of document
článek v časopisejournal article
Peer-reviewed
submittedVersion
Author
Janošek M.
Vyhnánek J.
Zikmund A.
Butvin P.
Butvinová B.
Rights
openAccessMetadata
Show full item recordAbstract
The effect of demagnetizing factor and manufacturing procedure on noise of the fluxgate probes, manufactured
from metallic glasses, is presented. The fluxgate probes were either tape-wound, or flat, wet-etched ringcore and
race-track geometries. When combining low demagnetizing factor and high enough cross-section of the probe, a
minimum in the noise dependence can be found. For 50 mm ringcores, in such minimum the noise value as low
as 7 pTRMS in 0.1-10 Hz frequency band was achieved, which is very low for an as-cast tape, and has not been
reported before.