JANOŠEK, M., et al. Effects of Core Dimensions and Manufacturing Procedure on Fluxgate Noise. Acta Physica Polonica A. 2014, 126(1), 104-105. ISSN 0587-4246. DOI 10.12693/APhysPolA.126.104.
The effect of demagnetizing factor and manufacturing procedure on noise of the fluxgate probes, manufactured
from metallic glasses, is presented. The fluxgate probes were either tape-wound, or flat, wet-etched ringcore and
race-track geometries. When combining low demagnetizing factor and high enough cross-section of the probe, a
minimum in the noise dependence can be found. For 50 mm ringcores, in such minimum the noise value as low
as 7 pTRMS in 0.1-10 Hz frequency band was achieved, which is very low for an as-cast tape, and has not been
reported before.