Zobrazit minimální záznam



dc.contributor.authorJanošek M.
dc.contributor.authorButta M.
dc.contributor.authorDressler M.
dc.contributor.authorSaunderson E.
dc.contributor.authorNovotný D.
dc.contributor.authorFourie C.
dc.date.accessioned2024-05-16T10:41:44Z
dc.date.available2024-05-16T10:41:44Z
dc.date.issued2020
dc.identifierV3S-334308
dc.identifier.citationJANOŠEK, M., et al. 1-pT noise fluxgate magnetometer for geomagnetic measurements and unshielded magnetocardiography. IEEE Transactions on Instrumentation and Measurement. 2020, 69(5), 2552-2560. ISSN 0018-9456. DOI 10.1109/TIM.2019.2949205.
dc.identifier.issn0018-9456 (print)
dc.identifier.issn1557-9662 (online)
dc.identifier.urihttp://hdl.handle.net/10467/114419
dc.description.abstractWe present the development of a low-noise, fundamental-mode, orthogonal fluxgate magnetometer with four amorphous, annealed ferromagnetic wires. The 1-Hz noise obtained in open-loop and closed-loop is as low as 0.75 and 1.5 pTrms/√Hz, respectively, with white noise level about 0.6 pTrms/√Hz. This is to our knowledge the lowest figure published for a fluxgate magnetometer so far. We compared the instrument performance to a low-noise observatory magnetometer when doing geomagnetic measurements and show that it is fully suitable for measurements at mHz frequencies, e.g. magnetotellurics. The magnetometer performance enables room-temperature, unshielded magnetocardiography.eng
dc.format.mimetypeapplication/pdf
dc.language.isoeng
dc.publisherIEEE
dc.relation.ispartofIEEE Transactions on Instrumentation and Measurement
dc.rightsCreative Commons Attribution-ShareAlike (CC BY-SA) 4.0
dc.rights.urihttp://creativecommons.org/licenses/by-sa/4.0/
dc.subjectmagnetometereng
dc.subjectnoiseeng
dc.subjectmagnetocardiographyeng
dc.subjectgeomagneticeng
dc.title1-pT noise fluxgate magnetometer for geomagnetic measurements and unshielded magnetocardiographyeng
dc.typečlánek v časopisecze
dc.typejournal articleeng
dc.identifier.doi10.1109/TIM.2019.2949205
dc.relation.projectidinfo:eu-repo/grantAgreement/Czech Science Foundation/GJ/GJ16-10591Y/CZ/Magnetic gradiometer based on fundamental mode orthogonal fluxgate/
dc.relation.projectidinfo:eu-repo/grantAgreement/EC/OPVVV/CZ.02.2.69%2F0.0%2F0.0%2F16_027%2F0008465/CZ/International Mobility of Researchers in CTU/
dc.rights.accessopenAccess
dc.identifier.wos000528560200026
dc.type.statusPeer-reviewed
dc.type.versionacceptedVersion
dc.identifier.scopus2-s2.0-85083230840


Soubory tohoto záznamu


Tento záznam se objevuje v následujících kolekcích

Zobrazit minimální záznam

Creative Commons Attribution-ShareAlike (CC BY-SA) 4.0
Kromě případů, kde je uvedeno jinak, licence tohoto záznamu je Creative Commons Attribution-ShareAlike (CC BY-SA) 4.0