On a static and dynamic calibration of thermochromic liquid crystals
Type of document
articlePeer-reviewed
publishedVersion
Author
Solnař, Stanislav
Medek, Jan
Suleiman, Abubakar Shola
Vyhlídal, Patrik
Rights
Creative Commons Attribution 4.0 International Licensehttp://creativecommons.org/licenses/by/4.0/
openAccess
Metadata
Show full item recordAbstract
The paper deals with a static and dynamic calibration of selected wide-range thermochromic liquid crystals (TLC - SolarDust 24C), which are very easy to ingest and can provide accurate information on the temperature distribution on different surfaces. Static calibration problems, such as TLC illumination angle or illumination intensity, are solved and conclusions are drawn for the application of such a measurement. Dynamic experiments also indicate a certain time delay (around 50 ms) of the applied TLC measurement layer, which is very important to know for dynamic experimental methods, but the error and inaccuracy of the experiment is too high to draw conclusions.
Collections
The following license files are associated with this item:
Except where otherwise noted, this item's license is described as Creative Commons Attribution 4.0 International License