ZIKMUND, A., et al. Precise Scalar Calibration of a Tri-Axial Braunbek Coil System. IEEE Transactions on Magnetics. 2015, 51(1), ISSN 0018-9464. DOI 10.1109/TMAG.2014.2357783.
A calibration of the well-defined Braunbek coil system was carried out using the scalar method. The whole measuring setup was designed to minimize the uncertainty of the scalar calibration procedure. The measurement time as well as the sampling ratio were adjusted to reduce the influence of the ambient magnetic field variation. We calibrated the coil sensitivity with the uncertainty of 30 ppm and orthogonality with the uncertainty <0.01°. The results were compared with a different technique.
eng
dc.format.mimetype
application/pdf
dc.language.iso
eng
dc.publisher
IEEE
dc.relation.ispartof
IEEE Transactions on Magnetics
dc.subject
Angle
eng
dc.subject
Braunbek system
eng
dc.subject
Calibration
eng
dc.subject
Coils
eng
dc.title
Precise Scalar Calibration of a Tri-Axial Braunbek Coil System
info:eu-repo/grantAgreement/Ministry of Education, Youth and Sports/ED/ED2.1.00%2F03.0091/CZ/University center for energy Efficient buildings (UCEEB)/UCEEB